This is an article for Ascott corrosion test chamber, The model from Ascott : CC450xp (CCT Chamber). It is the floor standing cyclic corrosion test chamber, Capacity is 450 Liters / 15.8 cu.ft. This CC450xp Ascott corrosion test chamber will pass through a 850mm wide door aperture for ease of installation.
As well as performing to the requirements of the traditional ASTM B117 based, continuous salt spray test standards. This CC450xp Ascott corrosion test chamber is also equipped with a control system to enable it to perform cyclic corrosion tests. These tests usually comprise of sequential periods of :
1. Salt spraying
2. Controlled humidity air drying
3. Condensation humidity (wetting)
The programmable control system fitted, enables the above conditions to be arranged in any sequence and permits other events to be programmed in between e.g. cold air blow. Also the duration of each event and number of times the whole cycle is to be automatically repeated, can be varied. Up to 10 multi-step cyclic programmes, 5 multi-step sub programs and 5 single step/single set point programs can be created and stored in the controller’ system memory for subsequent execution by the CC450xp Ascott cyclic corrosion test chamber.
Salt fog fallout rates: | Adjustable from 0.5 to 3.0ml per 80cm2 per hour | |
Air saturator temperature range: | Adjustable from ambient to +63℃ | |
Chamber temperature range: | Condensation humidity mode: | Adjustable from ambient to +55℃ /131℉ |
Salt spray mode (CC450xp CCT chamber) | Adjustable from ambient to +50℃/122℉ | |
Air drying and controlled humidity mode: | Adjustable from ambient to +70℃/158℉ | |
Chamber humidity range: | Condensation humidity mode: | Fixed at 95-100% RH |
Salt spray mode (CC450xp CCT chamber) | Fixed at 95-100% RH | |
Air drying/controlled humidity mode: | Adjustable (See Graph Below for Range) | |
Notes to the graph: | For CCT chambers without option ACC29, the minimum achievable temperature/humidity inside the cyclic corrosion test chamber (i.e.; the curved part of the graph) depends directly upon the ambient temperature/humidity of the air in the room in which the chamber is located. This part of the adjacent graph assumes room ambient conditions do not exceed 20℃/70℉ and 50% RH. | |
The CC450xp Ascott corrosion test chamber will operate in ambient conditions of up to 30℃/90℉ and 85% RH (non condensing). However, the minimum operating temperature/humidity that can be controlled inside the chamber, will be reduced accordingly. |
| Standard Accessories |
| Construction|CC450xp Ascott Corrosion Test Chamber |
| Operation and Control|CC450xp Ascott Corrosion Test Chamber |